Development and validation of a test for the measurement of methhemoglobin through the analysis of images with a smartphone - 2023

dc.contributor.advisorSaldaña Orejón, Italo Moisés
dc.contributor.authorLeon Trujillo, Bryan Estiff
dc.date.accessioned2024-08-28T16:29:06Z
dc.date.available2024-08-28T16:29:06Z
dc.date.issued2024-07-20
dc.description.abstractObjective: To develop and validate a test for measuring methemoglobin using image analysis with a smartphone. Materials and Methods: The design of this research is descriptive cross-sectional, correlational, quantitative, and prospective. It will consist of a set of arterial blood samples collected in lithium heparin tubes at the Carlos Lanfranco La Hoz Hospital in the biochemistry area from September 2023 to January 2024. Results: A linear relationship was found between sodium nitrite concentration and methemoglobin concentration obtained in the ABL-800 Flex gasometer, with an R² value of 0.9971. The values of the red channel showed an inversely proportional relationship with methemoglobin concentration. Image analysis was performed using the regression equation Y = -1.0036x + 132.44. According to the research, methemoglobin determination with a smartphone showed minimal bias compared to hemoglobin levels and methylene blue concentration. Conclusions: A simple and cost-effective method was developed for determining methemoglobinemia through measurement of the red channel values of RGB using a smartphone camera, with minimal interference from hemoglobin levels and methylene blue. This will be very useful in health centers with limited resources or in situations where methemoglobin concentrations cannot be measured rapidly.en_US
dc.description.abstractObjetivo: Desarrollar y validar un test para la medición de metahemoglobina mediante el análisis de imágenes con un Smartphone. Material y métodos: El diseño del presente trabajo de investigación es descriptivo transversal, de nivel correlativo, cuantitativo y prospectivo; estará conformada por un conjunto de muestras de sangre arterial extraídas en tubos con heparina de litio en el Hospital Carlos Lanfranco La Hoz en el área de bioquímica, en los meses de Setiembre del 2023a Enero del 2024. Resultados: Se evidencio una relación lineal entre la concentración de nitrito de sodio y la concentración de metahemoglobina obtenida en el gasómetro ABL-800 Flex, donde su valor de R2 fue igual a 0.9971. Así mismo, se observó que los valores del canal red presento una relación inversamente proporcional a la concentración de metahemoglobina, por ello, se realizó las mediciones por análisis de imagen mediante la ecuación de regresión logística Y= -1.0036x+132.44, así mismo, según la investigación la determinación de metahemoglobina con el celular presentó un sesgo menor respecto al nivel de hemoglobina y concentración de azul de metileno. Conclusiones: se desarrolló un método sencillo y económico de utilizar para la determinación de metahemoglobinemia, a través de la medición del canal red de valores RGB con la cámara de un Smartphone, obteniéndose además poca interferencia con los niveles de hemoglobina y azul de metileno. Lo cual será de mucha utilidad en centros de salud con recursos limitados, o en circunstancias donde las concentraciones de metahemoglobina no se pueden medir rápidamente.es_ES
dc.description.researchareaSalud y Bienestar
dc.formatapplication/pdf
dc.identifier.urihttps://hdl.handle.net/20.500.13053/11812
dc.language.isospaes_ES
dc.language.isospa
dc.publisherUniversidad Privada Norbert Wieneres_ES
dc.publisher.countryPEes_ES
dc.rightshttp://purl.org/coar/access_right/c_abf2
dc.rights.urihttps://creativecommons.org/licenses/by-nc-sa/4.0/
dc.sourceRepositorio Institucional de la Universidad Privada Norbert Wieneres_ES
dc.subjectMetahemoglobinaes_ES
dc.subjectMethemoglobines_ES
dc.subjectAnálisis espectrales_ES
dc.subjectSpectrum Analysises_ES
dc.subjectTeléfono inteligentees_ES
dc.subjectSmartphonees_ES
dc.subject.ocdehttps://purl.org/pe-repo/ocde/ford#2.06.02
dc.subject.odsODS 3: Salud y bienestar. Garantizar una vida sana y promover el bienestar de todos a todas las edades
dc.titleDevelopment and validation of a test for the measurement of methhemoglobin through the analysis of images with a smartphone - 2023en_US
dc.titleDesarrollo y validación de un test para la medición de metahemoglobina mediante el análisis de imágenes con un smartphone - 2023es_ES
dc.title.alternativeDevelopment and validation of a test for the measurement of methhemoglobin through the analysis of images with a smartphone - 2023es_ES
dc.typehttp://purl.org/coar/resource_type/c_7a1f
dc.type.versionhttp://purl.org/coar/version/c_970fb48d4fbd8a85
renati.advisor.cext 
renati.advisor.dni10042008
renati.advisor.orcidhttps://orcid.org/0000-0003-2389-7984
renati.advisor.pasaporte 
renati.author.cext 
renati.author.dni61178196
renati.author.pasaporte 
renati.discipline915126
renati.jurorAstete Medrano, Delia Jessica
renati.jurorQuintana Alfaro, Luis Alberto
renati.jurorValenzuela Martínez, Stefany Saragoza
renati.levelhttps://purl.org/pe-repo/renati/level#tituloProfesional
renati.typehttps://purl.org/pe-repo/renati/type#tesis
thesis.degree.disciplineTecnología Médica en Laboratorio Clínico y Anatomía Patológicaes_ES
thesis.degree.grantorUniversidad Privada Norbert Wiener. Facultad de Ciencias de la Saludes_ES
thesis.degree.levelTítulo Profesionales_ES
thesis.degree.nameLicenciado en Tecnología Médica en Laboratorio Clínico y Anatomía Patológicaes_ES

Archivos

Bloque original

Mostrando 1 - 3 de 3
Cargando...
Miniatura
Nombre:
T061_61178196_T.pdf
Tamaño:
2.26 MB
Formato:
Adobe Portable Document Format
Descripción:
Cargando...
Miniatura
Nombre:
FORM_61178196_T.pdf
Tamaño:
199.18 KB
Formato:
Adobe Portable Document Format
Descripción:
Cargando...
Miniatura
Nombre:
TURNITIN_61178196_T.pdf
Tamaño:
2.72 MB
Formato:
Adobe Portable Document Format
Descripción:

Bloque de licencias

Mostrando 1 - 1 de 1
Cargando...
Miniatura
Nombre:
license.txt
Tamaño:
1.71 KB
Formato:
Item-specific license agreed upon to submission
Descripción: